Important notes for Glovebox Circulation
Nov 08, 2024

Important notes for Glovebox Circulation

  • Set the gas cylinder pressure to 60-80 psi. Insufficient pressure may prevent the initiation or proper maintenance of circulation.  

  • When the antechamber is not in use, maintain a slight negative pressure and ensure both "Evacuation" and "Refill" are in the "off" position. 

  • Circulation cannot be initiated during the Purging or Regeneration processes. 

  • The circulation blower automatically adjusts its operating frequency based on moisture and oxygen levels, typically running at either 15 Hz or 30 Hz. 

  • If moisture and oxygen levels do not decrease after an extended period of circulation, the purifier may require regeneration, or you may need to contact a VTI service engineer for support. 